Rennex, B.
Title
Error Analysis for the National Bureau of Standards 1016 mm Guarded Hot Plate.
Coporate
National Bureau of Standards, Washington, DC
Sponsor
Department of Energy, Washington, DC
Report
NBSIR 83-2674, February 1983, 47 p.
Distribution
Available from National Technical Information Service
Contract
DE-AI01-76PR06010
Keywords
error analysis | guarded hot plate | thermal conductivity | thermal insulation | thermal resistance
Abstract
An error analysis is given for the 1-meter Guarded Hot Plate at the National Bureau of Standards. This apparatus is used to measure the thermal resistance of insulation materials. The individual contributions to uncertainty in thermal resistance are discussed in detail. The total uncertainty is estimated to be less than 0.5 percent at sample thicknesses up to 150 mm (6 inches) and less than 1 percent at a thickness of 300 mm (12 inches).