FireDOC Search

Author
Bryan, J. L. | DiNenno, P. J.
Title
Examination and Analysis of the Dynamics of the Human Behavior in the Fire Incidents at the University of Maryland Hospital on February 6, 1979.
Coporate
Maryland Univ., College Park
Sponsor
National Bureau of Standards, Gaithersburg, MD
Report
NBS GCR 80-234, March 31, 1979, 29 p.
Distribution
Available from National Technical Information Service
Contract
GRANT-79014
Keywords
fire departments | fire extinguishers | hospitals | nursing staff | patients
Abstract
This fire incident at the University of Maryland Hospital on February 6, 1979 was detected by a nurse at approximately 0840 hours. The nurse was alerted to the observation of flames in a microwave oven in the emergency room doctor's lounge, room G-1142, by a loud unusual noise. The nurse immediately reported the fire at the emergency room treatment area nurses station. The hospital emergency procedures were initiated with a phone call to the University Communication Center and activation of the building local alarm system. The fire in the microwave oven was extinguished by the nurse using a 10 pound, listed carbon dioxide extinguisher, rated 5 B, C. Following extinguishment a light haze of smoke was confined to the doctor's lounge. One patient in an adjacent emergency room treatment area was evacuated as a precautionary action. The fifteen story North Hospital Building of fire resistive construction was approximately five years old. At the time of the fire incident the twenty-five patient emergency room treatment area had ten patients. The Baltimore City Fire Department responded and verified extinguishment.